If dynamic forces are applied to equipment components that support or support patients during normal use (such as sitting, standing, patient handling, etc.), unacceptable risks must not occur.
Durability should consider the most unfavorable position of the bed components that support or support the patient during normal use. Verify compliance through the following tests: Test conditions:
Place the manufacturer's worst-case scenario mattress on a flat bed. If the height is adjustable, adjust the height to the worst-case position. Use a load pad (Figure 201.104) at position A in Figure 201.116, on the weakest side.
Force range: 0-2000N can be set
Using a stepper motor to drive the screw lifting mechanism for pressurization, sustainable or cyclic loading
Frequency: 6-15 times/min adjustable
Loading speed: 0.01-20mm/s can be set
Cycle times: 1-999999 can be set. After reaching the set time, the automatic cycle test function will be activated. When the test reaches the user set number of times, the device will automatically shut down and issue an alarm.
Specification of dynamic load test head: The load pad (see Figure 201.104) is a rigid circular object with a diameter of 355mm ± 5mm. Its surface has a convex spherical curvature with a radius of 800mm, and the front edge radius is 20mm.