201.9.8.3.3.3 "Sidebar strength and latch reliability" shall be designed and manufactured in accordance with the requirements of section 201.9.8.3.3.3 "Sidebar strength and latch reliability" in YY9706.252-2021 "Medical Electrical Equipment Part 2-52: Special Requirements for Basic Safety and Basic Performance of Medical Beds". Suitable for testing the load, guardrail push-pull strength test, and dynamic load fatigue test of medical beds, elderly care beds, traditional Chinese medicine therapy beds, and other products. Test the strength of the simulated bed under load conditions, as well as the push-pull strength of the guardrail and the dynamic strength of the tension rod. Compliance will be verified through the following tests, with the sidebar located in the upper position: a) Lateral force repeated action cycle test. Apply a force of 100 N perpendicular to the top center of the sidebar in the direction indicated by E or F in Figure 201.117. negative direction. Repeat 3000 times; b) Longitudinal force repeated action cycle test. Apply a force of 100 N along the longitudinal direction of the edge fence as shown in C or D of Figure 201.117. negative direction. Repeat 3000 times; c) Vertical force repeated action cyclic test. As shown in B of Figure 201.117, apply a force of 100N along the vertical direction of the sidebar to its top. Repeat 3000 times. A direction does not require testing; d) After completing a), b), and c), apply a static load in the direction of Figure 201.117 along the worst-case position. The sidebar must not become unlocked or generate other unacceptable risks. The range of force values is 0-1000N, and the accuracy of force values is ± 0.5%; Drive mode: motor-driven; Action range: 0~100mm can be set; Loading speed: 0.01-20mm/s, adjustable holding time: 1-99 (s) can be set arbitrarily, and the number of automatic pressure relief tests after the time is up: 999999 can be set. After reaching the set time, the automatic cyclic testing function will be activated, and the equipment will automatically stop and sound an alarm after reaching the user set number of tests.