Supporting chemical filter testing bench

Product Mode: Fixed mode of products: Test Speed: Number of tests: Equipment size: Description: This module is a specialized expansion unit for automotive filters in the chemical filter test bench, which meets the requirements of ISO 11155-1 and ISO 11155-2 for testing gas-phase pollutants in road vehicle cabin air filters. It is specifically designed for automotive air conditioning chemical filters (activated carbon composite filter, molecular sieve deodorization filter, composite VOC removal filter) and can be directly integrated into the main test bench and the original 0.1-90ppm air distribution system

Main technical parameters


Applicable samples


Automotive cabin panel air conditioning filter, composite activated carbon filter, PM2.5+gas-phase composite filter with chemical adsorption layer


Core testing items (ISO11155 requirements)


1. Initial filtration efficiency of target gas-phase pollutants (upstream and downstream synchronous sampling detection)


2. Characteristics of filter resistance and pressure difference under different air volumes


3. Dynamic adsorption breakthrough curve and saturated adsorption capacity


4. Verification of the influence of temperature and humidity on adsorption efficiency under simulated real vehicle operating conditions


5. Intermittent on-off test (simulating vehicle start stop cycle intake)


6. Particle synchronous testing (optional, ISO11155 also includes particle segment requirements)


Module composition


1. Standard fixture for car filter cartridge

Adapt to mainstream car filter sizes in the market (210 × 180, 240 × 190, 280 × 220, etc.), seal the frame to prevent bypass leakage (ISO11155 strictly controls bypass leakage); Support quick sample changes.


2. Standard surface wind speed control subunit

Match the test surface wind speed specified in ISO11155: 0.25 m/s (standard operating condition), which can be accurately adjusted between 0.1~0.8m/s, with closed-loop air volume control.


3. Precise control section for temperature and humidity under working conditions

According to standard requirements: 23 ℃± 2 ℃, 50% RH ± 5% RH reference condition; It can also be extended to high temperature and high humidity (simulating the interior environment of a car in summer).


4. Dual synchronous sampling manifold (ISO sampling structure)

Upstream and downstream symmetrical sampling probes, isokinetic sampling, to avoid concentration sampling deviation; All pipelines are made of PTFE low adsorption material to prevent VOC/acidic gases from being adsorbed by the pipelines and causing false high data.


5. Operating cycle control unit

Programmable intermittent intake, start stop cycle, simulate intermittent opening of vehicle air conditioning, and conduct accelerated aging/life penetration tests.


6. Standard algorithm software package (ISO11155 specific)

Built in ISO11155 calculation formula, automatically outputting report templates; Automatically distinguish between particle filtration efficiency and gas-phase chemical adsorption efficiency.


Key technical parameters


Test surface wind speed: 0.10-0.80 m/s, closed-loop control, accuracy ± 0.02 m/s


Standard test surface wind speed: 0.25 m/s (as specified in ISO11155)


• Suitable sample size: customized according to needs, compatible with mainstream car air conditioning filters


Test gases: toluene, formaldehyde, NO ₂, SO ₂, NH ∝ (target gas can be added as needed)


• Concentration using host range: 0.1-90 ppm


• Differential pressure range: 0~1000Pa (vehicle filter resistance is generally low)


• Temperature control: 18-40 ℃; Humidity control: 30% to 80% RH


Leak verification: Comes with a bypass leak self check program


Test Procedure (ISO11155 Standard Procedure)


Install the tooling into the vehicle filter, seal it, set the standard wind speed, temperature and humidity, preprocess the clean air duct of the

 host, stabilize the target pollutant concentration in the gas distribution, collect the upstream data, continuously pass the gas through the 

filter, continuously sample and record the concentration downstream, record the pressure difference and concentration curve in real time, c

alculate the initial efficiency of the software according to the ISO formula, draw the penetration curve, automatically stop the test when 

the penetration threshold is reached, and generate a standard report.



c125804ae60fb81d9ea3f8630cd3c094.jpg

8a3951fc2708b565b8d63ffb64b547fc.jpg

79fc579b17bd9e62ce6cb5af9784795f.jpg

  • 官网二维码

    官网二维码

  • 国际网站

    国际网站

  • 阿里巴巴

    阿里巴巴

广州市艾利讯电子科技有限公司
联系电话:Wechat /Facebook / Whatsapp
+8618928987982 020-39921845(伍经理)
公司电话:020-39921845
企业邮箱:qx668798@163.com / allieson@163.com
公司官网:www.allison-tm.com
工厂地址:广州市番禺区石基镇旦岗西街22号
GUANGZHOU ALLISON ELECTRONIC TECHNOLOGYCO.,LTD.
Add: No. 22, Xijie West Street, Danggang Town, Shi Ji Town, Panyu District, Guangzhou City, Gudong Province, China
Tell / Wechat / Facebook / Whatsapp:+8615915832210 (Allie)
Tell:+86 020-39921845
Mail: qx668798@163.com / allieson@163.com