Applicable samples
Automotive cabin panel air conditioning filter, composite activated carbon filter, PM2.5+gas-phase composite filter with chemical adsorption layer
Core testing items (ISO11155 requirements)
1. Initial filtration efficiency of target gas-phase pollutants (upstream and downstream synchronous sampling detection)
2. Characteristics of filter resistance and pressure difference under different air volumes
3. Dynamic adsorption breakthrough curve and saturated adsorption capacity
4. Verification of the influence of temperature and humidity on adsorption efficiency under simulated real vehicle operating conditions
5. Intermittent on-off test (simulating vehicle start stop cycle intake)
6. Particle synchronous testing (optional, ISO11155 also includes particle segment requirements)
Module composition
1. Standard fixture for car filter cartridge
Adapt to mainstream car filter sizes in the market (210 × 180, 240 × 190, 280 × 220, etc.), seal the frame to prevent bypass leakage (ISO11155 strictly controls bypass leakage); Support quick sample changes.
2. Standard surface wind speed control subunit
Match the test surface wind speed specified in ISO11155: 0.25 m/s (standard operating condition), which can be accurately adjusted between 0.1~0.8m/s, with closed-loop air volume control.
3. Precise control section for temperature and humidity under working conditions
According to standard requirements: 23 ℃± 2 ℃, 50% RH ± 5% RH reference condition; It can also be extended to high temperature and high humidity (simulating the interior environment of a car in summer).
4. Dual synchronous sampling manifold (ISO sampling structure)
Upstream and downstream symmetrical sampling probes, isokinetic sampling, to avoid concentration sampling deviation; All pipelines are made of PTFE low adsorption material to prevent VOC/acidic gases from being adsorbed by the pipelines and causing false high data.
5. Operating cycle control unit
Programmable intermittent intake, start stop cycle, simulate intermittent opening of vehicle air conditioning, and conduct accelerated aging/life penetration tests.
6. Standard algorithm software package (ISO11155 specific)
Built in ISO11155 calculation formula, automatically outputting report templates; Automatically distinguish between particle filtration efficiency and gas-phase chemical adsorption efficiency.
Key technical parameters
Test surface wind speed: 0.10-0.80 m/s, closed-loop control, accuracy ± 0.02 m/s
Standard test surface wind speed: 0.25 m/s (as specified in ISO11155)
• Suitable sample size: customized according to needs, compatible with mainstream car air conditioning filters
Test gases: toluene, formaldehyde, NO ₂, SO ₂, NH ∝ (target gas can be added as needed)
• Concentration using host range: 0.1-90 ppm
• Differential pressure range: 0~1000Pa (vehicle filter resistance is generally low)
• Temperature control: 18-40 ℃; Humidity control: 30% to 80% RH
Leak verification: Comes with a bypass leak self check program
Test Procedure (ISO11155 Standard Procedure)
Install the tooling into the vehicle filter, seal it, set the standard wind speed, temperature and humidity, preprocess the clean air duct of the
host, stabilize the target pollutant concentration in the gas distribution, collect the upstream data, continuously pass the gas through the
filter, continuously sample and record the concentration downstream, record the pressure difference and concentration curve in real time, c
alculate the initial efficiency of the software according to the ISO formula, draw the penetration curve, automatically stop the test when
the penetration threshold is reached, and generate a standard report.