Provide stable and reliable environmental conditions for 1-1 refrigerator type testing and development. The interior of the testing room can accommodate 6 refrigerators for various standard operating conditions and simulated usage conditions for type testing.
The 1-2 laboratory complies with the requirements of sections 6.1.1 to 6.1.4 of national standards GB8059.1 to 3-1995, sections 5.1 to 5.1.2 of GB8059.4-93, and meets the requirements of the new version of GB/T8059 (draft for approval). The system is fully automatic for working condition control, automatic measurement and recording, automatic printing of test reports, and can export data through USB connectors for analysis of test results.
The 1-3 laboratory can provide the following simulated testing environments:
The laboratory can provide the following simulated testing environment for the needs of cold and warm box testing:
Temperature regulation range: RT+10 ℃~50 ℃;
The external dimensions of the chamber (width x depth x height) shall not exceed (4500 x 6000 x 3200) mm
Thermal insulation: polyurethane foam (PUF board)
Humidity: 45%~75% RH ± 3% (humidity controlled between 16~43 ℃)
Temperature deviation: ± 0.5 ℃;
Test angle (fireproof board): 1300 × 1300mm, made of wood or wooden material, painted matte black
The opening height must be at least 2200 millimeters, and the opening width must be at least 1500 millimeters
Humidity deviation: The displayed value of the digital temperature and humidity control instrument is ± 2% RH within the dry bulb range of 15-40 ℃;
Temperature gradient: less than 1.0 ℃/m above 300mm above the ground in the laboratory;
Environmental wind speed: measured according to the national standard location requirements, with a maximum value of ≤ 0.25m/s measured in any direction;
Temperature distribution accuracy in the laboratory:
When there is no load: ± 0.5 ℃ in the inner space 0.5m away from the wall, floor, and ceiling;
When under load: the vertical temperature gradient is less than 1 ℃/m
The measurement range of air flow velocity must not be less than (0.1... 5) m/s
Internal lighting: 6 sets
Storage room: 2MX2MX2.5M high (optional)
Set a slope at the threshold to facilitate loading the test product into the test chamber
The deviation of the horizontal temperature difference at a distance of 1m from the ground and 350mm from the center of the front and side of the refrigerator is ± 0.5 ℃.