This experimental setup consists of a laboratory, operating condition control equipment, etc. The testing room adopts an assembled storage plate structure, with an external polyurethane colored plastic storage plate made of material, with a thickness of 100mm. A 3mm thick stainless steel plate is added to the middle aisle, and 1.5mm stainless steel orifice plates are set below the testing stations on both sides. The top of the laboratory adopts aluminum alloy orifice plates.
The environmental control of the 1-2 laboratory body adopts top air supply and bottom air return methods to ensure the uniformity of temperature and humidity parameters. Internally equipped with an air conditioning unit, two independent power switches for each workstation, and a workstation testing bench. (Based on the customer's existing environmental warehouse)
Equipped with a testing instrument cabinet, computer operation console, main control computer, PLC control system, high-precision data acquisition system, high-precision temperature and humidity controller, etc. The laboratory operating condition control equipment and control cabinet are installed nearby on one side of the laboratory body, and are separated from the testing signal line for wiring to reduce interference with measurement parameters. (Based on the customer's existing environmental warehouse)
The 1-3 laboratory adopts an assembled storage plate structure, with an external polyurethane colored plastic storage plate material and a thickness of 100mm. A 3mm thick stainless steel plate is added to the middle aisle, and 1.5mm stainless steel orifice plates are set below the test bench on both sides for return air. The top of the laboratory is equipped with aluminum alloy orifice plates for air supply. Insulated door and observation window (based on the customer's existing environmental warehouse)
1-4 adopts a single opening insulated door, with a net size of 1500mm × 2200mm for the door frame (W × H). (Based on the customer's existing environmental warehouse)
A double-layer vacuum glass observation window with dimensions (W × H) of 400mm × 300mm is embedded and installed on the insulation door 1-5, used to observe the internal conditions of the test chamber during the test process and avoid the impact of frequent door opening on the test conditions.
Air circulation system
1-6 adopts a laminar air circulation method with suspended ceiling air supply and floor return air. Both the supply and return air outlets are designed with orifice plate structures, ensuring a balanced flow direction of indoor airflow from top to bottom. This maximizes the uniformity of indoor temperature and humidity parameters, and guarantees that the wind speed at all locations is less than the 0.25m/s specified in the testing standards. (Based on the customer's existing environmental warehouse)
The 1-7 testing stations are evenly distributed on both sides of the laboratory aisle, and are 300mm above the ground. The bottom of the station is a stainless steel perforated plate for return air. Each station is separated by black matte partitions with a width of 300mm.