1. Power supply: 220v50Hz/380v50Hz
2. Temperature range: -100℃---300℃
3. Temperature control accuracy: ±0.05%;
4. Temperature uniformity ≤0.01℃
5. Resistance measurement range, 0.001-1MΩ, accuracy 0.001mΩ
6. The tooling is suitable for installing EH/SM series terminal thermistors (see samples for details)
7. Cooling method: compressor refrigeration cooling
8. Medium: Silicone oil (-100-300℃)/alcohol
9. The lowest temperature rise rate is 0.5℃/min
10. Box body packing medium specification: 35cm*35cm
11. The software supports automatic data saving work
12. The equipment has leakage protection and undervoltage protection safety devices
13. Exportable report type data
14. Support multi-temperature point setting to automatically complete the operation (for example: 0℃-25℃-50℃-85℃-100℃-125℃-150℃)
15. The test temperature points in the automatic mode can be set at will. At least 10 temperature points can be set in a test process, and can be set from negative temperature to positive temperature, positive temperature to negative temperature, positive temperature to negative temperature and then Negative temperature and other programs.